Batch #6 (STEP1)
 
Analysis Summary Statistics Summary


  • 480 nm quantum efficiency


The mean value for the whole batch is about 18.3%, taking into account only the PMTs which fulfill the specifications in the calculations. If we compute the mean value for each sub-batch (250 PMTS each), one of them is just below than 18% as required. We have noticed this batch has a mean value slightly lower than the others. Besides, all tubes satisfy the criteria defined on the quantum efficiency (that must be greater than 15%).


The distribution of the 480 nm efficiency is reported on figure (1). The rejected PMTs (all) are quoted in red. As for previous batches, we consider that this criterion is ok.

 Next
Batch #6
1250 PMTS
(5 sub-batches)
tested by
  1. Clermont Ferrand
    (250 PMTS)
  2. Urbana
    (250 PMTS)
  3. Uta
    (250 PMTS)
  4. Dubna
    (250 PMTS)
  5. Valencia
    (250 PMTS)